Menu

The new NanoVIP® is a diverse, fully automated system for fluorescence in situ hybridization (FISH), in situ hybridization (ISH) and miRNA ISH. Its reliable automation combiningwith eXACT™ temperature modules and liquid level sensors for accurate liquid handling ensuresrobust and reproducible results.

The All-in-One system can run 10 different protocols simultaneously, making new protocol optimization easier. This instrument can simplify a complex 33-step manual FISH protocol and reduce it to 3simplesteps: load slides, select protocol, and view completed slides. The NanoVIP® software is an intuitive easy-to-use software whichcreates and stores user-defined protocols.With quality engineered hardware, unparalleled software capabilities, and one of the smallest footprints on the market, the NanoVIP remains an ideal choice for both clinical and research laboratories.

NanoVIP® Overview

Automation Full (baking through coverslip)
Slide capacity 10 slides
Reagent capacity 24 reagent vials
Reagent vial capacity 10 mL
Bulk reagent carboy & capacity 5 x 1 L
Waste container & capacity 1 x 4 L
Slide sensing Open
Reagent sensing Open
Reagent dispense volumes 10-180 µL
Reagent dispenser Pipette tips
Slide temperature range 25-105 °C
Configuration Table mounted
Protocols Open template
Operating temperature 15-30 °C
Operating humidity 15%-55% RH
Operating voltage 115/230 V AC, 60/50 Hz, 550 W
Dimensions (D/W/H) 20″ x 30″ x 20″
Weight 106 lb/48 Kg

Features and Benefits

  • 10 independent thermal cyclers positions –rapid heating and cooling (from ambient to 95 ºC in less than 2 minutes)
  • eXACT™ temperature control on every slide (±1.0 ºC)
  • 10 independent hybridizer stations for running up to ten protocols simultaneously
  • Removable slide racks for ease of operation
  • Onboard oil sealing and coverslipping to eliminate rubber cement and manual manipulation
  • Onboard washing with waste drainage system
  • Audio and visual alerts and on-screen color display for run status
  • Customizable protocols and built-in factory protocols
  • User-friendly software with retrievable experimental data
  • Covers all 3 major nucleic acid analysis methods: FISH, ISH, and in situPCR
  • Manual application of probes and other micro-reagents for cost saving
  • Run 10 different protocols simultaneously
  • User-friendly software with detailed report generation

 

NanoVIP® Video